Calibration of a Liquid Xenon Detector with Kr-83m

Kastens, L. W., Cahn, S. B., Manzur, A. et al.

Understand

We report the preparation of a Kr-83m source and its subsequent use in calibrating a liquid xenon detector.

  • Kr-83m atoms were produced through the decay of Rb-83 atoms trapped in zeolite molecular sieve and were then introduced into liquid xenon.
  • Decaying Kr-83m nuclei were detected through liquid xenon scintillation.
  • Conversion electrons with energies of 9.4 keV and 32.1 keV from the decay of Kr-83m were both observed.